and push down on the end with the pins. Make sure the cell assembly is flush with the lamp window.
4) Place the spacer around the assembly.
5) Place the filter media over the Cell Assembly centered on the top of the sensor. Make sure the filters are installed in
the correct order. Filter Media #2 first, then Filter Media #1 on top, with the shiny side up.
6) Align the Cap Key with the notch on the housing. Starting at the side opposite the notch, press down until the
Filter Cap snaps on to the housing. If the Cap Key is incorrectly aligned there will be a noticable buldge on the side of
the Cap.
PI-600 Toxic Gas Sensors PG.29